Turntable
TTT-510
●
Chemical resistant cleaning reservoir
●
Various electrode cleaning methods
※Optional air pump, reagent pump and others are
required for above cleaning methods except for
pure water shower.
●
Electrode holding tank available
for overnight operation
After the final measurement, the electrode is stored
in a storage tank filled with pure water, so there are
no concerns that the electrode will dry out.
Automatic Sampler
AST-3210
Sample measuring principle
Three-way stopvalve switching measuring tube
Sample volume
About 0.2
─
20 mL (fixed loop volume [measuring tube])
Measurement repeatability
0.5% or less for 10 mL sample as CV value
Measurement vessel
Material : hard glass, Number of electrode : Max. 4
Max. volume : approx. 150 mL, Number of titration nozzle : Max. 4
Min. required volume : about 100 mL
Alarm indication Pure water empty, Waste water full, System abnormal
Tanks for waste water &
pure water
20 L plastic tank
Power source AC line 50/60Hz
Power consumption Approx. 100VA
External dimensions & Weight
Approx. W565 mm × H635 mm × D480 mm, approx. 45 kg
*
The attachment cable for the titrator, trunk cable for electrode and others are sold
separately based on the model designation
Data Acquisition Software 701
7075650K
Capable of importing data to PC in the CSV format
using USB memory, LAN or RS-232.
Automatic Burette
ABT-7
Magnetic Stirrer
ST-7
External printer
EPS-R
Specifications
Display unit LCD display
Operating keys/switch
Flat keyboard
Number of samples
/beakers used
12 samples 200 mL, 300mL tall beaker
18 samples 100 mL tall beaker (Manufacturer's designation)
36 samples 30 mL, 50 mL tall beaker (Manufacture's designation)
60 samples/(100 samples) 20 mL cell (Manufacturer's designation)
Cleaning mode
①
Pure water shower → Air blow
②
Pure water bubbling → Pure water shower → Air blow
③
Liquid chemical shower → Pure water shower → Air blow
④
Liquid chemical bubbling → Pure water shower → Air blow
⑤
Liquid chemical bubbling → Pure water bubbling → Pure water shower → Air blow
⑥
Liquid chemical bubbling → Liquid chemical shower → Air blow
⑦
Liquid chemical shower → Air blow
⑧
Liquid chemical bubbling → Pure water bubbling → Liquid chemical shower → Air blow
Note : ・Air blow cleaning is effective when the optional air pump is attached.
・Liquid chemical cleaning is effective when the optional liquid chemical pump is attached.
・
Bubbling is effective when the optional bubbling pump and the waste water valve are attached.
Cleaning time 1
─
9,999 seconds (Setting possible)
Air blow frequency 1
─
9 times (Setting possible)
End detection Using endpin or setting final sample number by key operation
Cleaning tank material
Teflon, Daiflon
Electrode storage tank
Standard
Stirring time before
measurement
0
─
9,999 seconds
Waiting time before
measurement
0
─
9,999 seconds
Alarm
Cleaning water 1 (Pure water) Empty, Cleaning water 2 (Chemical) Empty, Waste water Full
Ambient temperature
5
─
40℃
Power source AC100V
─
240V 50/60Hz
Power consumption
Max. approx. 55VA (without optional pump & valve)
Max. approx. 175VA (with optional pump & valve)
External dimensions
Main unit : Approx. W380 mm × H392 mm × D622 mm
Maximum size when moving : Approx. W 415mm (with table plate) × H518 mm × D622 mm
Weight Approx. 18kg (not including table and electrodes)
Standard Accessories
Table (designated for 12, 18, 36 and 60 samples)
Electrode holder (designated samples / use)
Pure water tank (10 L, with float switch)
Waste water tank (10 L, with float switch)
Pure water tube
Waste water tube
End pin
Polyethylene Beaker (200 mL)
Instruction Manual
*
The attachment cable for the titrator is sold separately
based on the model designation.
Options
Description Code
Connection cable (2m) 7075670K
Air pump P000023
Liquid chemical pump P000022
Bubbling pump 0FA00001
Waste water valve (pinch valve) 0QA00001
Waste water valve (solenoid valve) 0QA00002